IceCure
™
Medical Ltd.
DSR3200000 rev. G
ProSense
™
Confidential
User Manual
89
Return to first page U.S User Manual
2.4 mm / 13G
134
mm
Trocar
Elliptic
14.5 mm
FAP7800000
•
Cryoprobe and matching introducer are marked in the same color.
Introducer
Shaft Length
Description
Tip Shape
Part Number
115 mm
Fit to needle diameter 13G (FAP7800000)
Trocar
FAC9000000
122 mm
Fit to needle diameter 10G (FAP7200000)
Trocar
FAC9100000
167 mm
Fit to needle diameter 10G
(FAP7400000 and FAP7410000)
Trocar
FAC9200000
*IceCure Medical systems, needles, introducers, cryoprobes and accessories may not be available in all
regions.
Figure 79: Introducer shaft length
The order in which you follow these steps is essential for maintaining sterility and
patient safety.
First,
insert the cryoprobe tip into the target tissue. Be aware of the markings on the
cryoprobe: the wide mark closest to the tip is the safety mark. In percutaneous
procedures
it must be completely
inside the tissue
to avoid skin burns. The rest of the
marks indicate depth of cryoprobe insertion: each mark equals one centimeter with
distinctive markings at 5 and 10 cm as shown in Fig. below.
Figure 80: illustration of FAP7200000 cryoprobe Markings on the cryoprobe.
Main marks indicate safety zone, 5 cm (slightly thick third mark from the left) and 10 cm
(double mark in circle).
Once you have verified the cryoprobe is located in the correct place, you may begin
freezing.
5.3.3
Freeze Cycle
Selecting the type of treatment, Manual or Automatic, is discussed in section 5.3.4.1 and
5.3.4.2.