Last change 04/2016 E-P – AL/TW/PW | Subject to technical modifications
Failure
Reason
Remedy
Dosing capacity of
pump not exact
Dosing head not de-aired
totally
Repeat de-airing
Out gassing material
Check installation
Valves partly contaminated or
crusted
Clean valves
False monitoring of dosing
capacity
Calibrate
Counter pressure fluctuations Installation of a pressure hold
valve and pulsation damper
(muffler)
Intake height fluctuations
Keep intake level constant
Pre-pressure higher than
counter pressure
Installation of pressure hold valve
Leaking or porous intake line
or pressure pipe
Replace intake line or pressure
pipe
Components coming in
contact with the material not
resistant
Replace by resistant materials
Worn dosing membrane
(initial cracks)
Replace membrane
Modification of dosing
material (density, viscosity)
Check concentration; possibly
install an agitator
Loud creaking noise
Resonance noises of the step
index motor may be audible
short-term at dosing
efficiency <10% of maximum
pump dosing capacity.
The resonance noise of the index
motor is no failure at all.
Liquid penetrates at
flange between pump
and dosing head
Faulty membrane
Replace membrane
Call our service personnel!
Содержание HydroDOS HD550G
Страница 19: ...I 19 103 Last change 04 2016 E P AL TW PW Subject to technical modifications 2 2 Installation schematics...
Страница 39: ...I 39 103 Last change 04 2016 E P AL TW PW Subject to technical modifications...
Страница 43: ...I 43 103 Last change 04 2016 E P AL TW PW Subject to technical modifications Calibration...
Страница 98: ...I 98 103 Last change 04 2016 E P AL TW PW Subject to technical modifications Technical documents...
Страница 102: ...I 102 103 Last change 04 2016 E P AL TW PW Subject to technical modifications 4 Installation drawing...