--------------------------------------------------------Auto Ref/Keratometer HRK-7000/7000A 87
②
Messages during measurement
Message
Cause
Method of Settlement
TRY AGAIN
Refer to page 15
Refer to page 15
Objective glass in the measurement
window is polluted
Clean the glass
+ OUT
Sphere of examinee’s eye exceeds
+22D
Impossible to
measure
Curvature radius of examinee’s eye
exceeds 10.2mm
Object lens within measurement
window is polluted
Clean the glass
- OUT
Sphere of examinee’s eye exceeds -
22D
Impossible to
measure
Curvature radius of examinee’s eye is
less than 5.0mm
Objective glass in measurement
window is polluted
Clean the glass
C OUT
Astigmatism of examinee’s eye
exceeds 10D
Impossible to
Measure
Corneal astigmatism of examinee’s
eye exceed 15D
Object lens within measurement
window is polluted
Clean the glass