--------------------------------------------------------Auto Ref/Keratometer HRK-7000/7000A 57
8.7.
Retro-ILLUM Measurement Mode (Retro-ILLUM Mode)
Retro-Illum measurement mode is the measuring function to use usefully in the
following cases.
1.
It is to examine the crystalline lens of patient who has the severe symptom of
cataract or undergoes it, or to measure its refractive power.
-
Examine the degree of opacity of crystalline lens with the shape of light reflected
from retina while changing intensity of light shed on the eye.
-
In case that the crystalline lens is not much in opacity, it is possible to measure
the sight refractive power of eye as well while observing the shape reflected
from retina.
2.
In case there are scratches on retina, observe the scratches: or observe whether
or not the penetration of light into IOL is uniform after the implantation surgery of
IOL.
8.7.1.
Adjustment of Array and Focus
①
Perform the adjustment of array and focus according to procedure 1, 2 of section
8.1.1.
②
As pushing ILLUM button after pushing FUNC button while selecting Ret-Illum
mode, the [Observe] screen below is to appear.
Содержание HRK-7000
Страница 1: ... Auto Ref Keratometer HRK 7000 7000A 1 User s Manual Auto Ref Keratometer HRK 7000 7000A ...
Страница 43: ... Auto Ref Keratometer HRK 7000 7000A 43 Figure 22 Example of Print ...
Страница 47: ... Auto Ref Keratometer HRK 7000 7000A 47 Figure 24 Example of Print ...
Страница 53: ... Auto Ref Keratometer HRK 7000 7000A 53 Figure 26 Screen indicating KER P Mode ...
Страница 74: ...74 Auto Ref Keratometer HRK 7000 7000A Figure 41 Display of VIRTUAL COMPARISON Mode ...
Страница 84: ...84 Auto Ref Keratometer HRK 7000 7000A ...
Страница 97: ... Auto Ref Keratometer HRK 7000 7000A 97 ...