62 Auto Ref/Keratometer HRK-7000/7000A --------------------------------------------------------
-
As pushing Mode button in [Observe] window, it is to be changed to [Measure]
window. At this time, as pushing Mode button again, [Observe] window is to
return. Position the bright dot which indicates LED light to shed on the eye so
that it can avoid the part of opacity of pupil by using the operation lever, and
stop the image and the measured value by pushing the measurement button
after well adjusting the focus of image appeared on the screen. If the stopped
image is not satisfactory, stop the image again after returning to [Measure]
window by pushing EXIT button.
NOTE
The opacity of crystalline lens caused by cataract can lead in errors of measured
value while causing the aberration by the decenteredness.
8.7.3.
Storage
If you want to store the stopped image in memory, push the measure button. You can
store max of two (2) images for each eye. If you want to return to [Observe] or[Measure]
window, please push EXIT button.
8.7.4.
Examination on the other eye
Perform the examination on the other eye and the storage of its image by the same way.
Содержание HRK-7000
Страница 1: ... Auto Ref Keratometer HRK 7000 7000A 1 User s Manual Auto Ref Keratometer HRK 7000 7000A ...
Страница 43: ... Auto Ref Keratometer HRK 7000 7000A 43 Figure 22 Example of Print ...
Страница 47: ... Auto Ref Keratometer HRK 7000 7000A 47 Figure 24 Example of Print ...
Страница 53: ... Auto Ref Keratometer HRK 7000 7000A 53 Figure 26 Screen indicating KER P Mode ...
Страница 74: ...74 Auto Ref Keratometer HRK 7000 7000A Figure 41 Display of VIRTUAL COMPARISON Mode ...
Страница 84: ...84 Auto Ref Keratometer HRK 7000 7000A ...
Страница 97: ... Auto Ref Keratometer HRK 7000 7000A 97 ...