56 Auto Ref/Keratometer HRK-7000/7000A --------------------------------------------------------
8.6.
Intraocular Lens (IOL) Measurement Mode (IOL Mode)
The cases that light is reflected from the surface of IOL, that the crystalline lens is in
opacity just like a cataract patient , or that the radius of pupil is very small can cause the
error in the measured value of refractive power. In these cases, measure it by pushing
IOL button.
In case of severe cataract, measure or observe it in Retro-Illum mode (refer to section
8.7).
1.
Selection of REF Mode or K&R Mode
-
Keep pushing MODE button until “REF“ or ”K&R” is to be indicated on the right
upper side of screen.
2.
Perform the adjustment of array and focus according to procedure 1, 2 of section
8.1.1.
3.
Selection of IOL Mode
-
Push IOL button (indicated as “REF-I” , “K&R-I”).
4.
Measurement
-
Push the measurement button.
-
As you keep pushing the measurement button, the measurement is to be
performed consecutively.
NOTE
Decenteredness and distortedness of eye implanted with IOL, or being right after
surgery can cause the error to the measured values as there is a case of deformation of
iris.
Содержание HRK-7000
Страница 1: ... Auto Ref Keratometer HRK 7000 7000A 1 User s Manual Auto Ref Keratometer HRK 7000 7000A ...
Страница 43: ... Auto Ref Keratometer HRK 7000 7000A 43 Figure 22 Example of Print ...
Страница 47: ... Auto Ref Keratometer HRK 7000 7000A 47 Figure 24 Example of Print ...
Страница 53: ... Auto Ref Keratometer HRK 7000 7000A 53 Figure 26 Screen indicating KER P Mode ...
Страница 74: ...74 Auto Ref Keratometer HRK 7000 7000A Figure 41 Display of VIRTUAL COMPARISON Mode ...
Страница 84: ...84 Auto Ref Keratometer HRK 7000 7000A ...
Страница 97: ... Auto Ref Keratometer HRK 7000 7000A 97 ...