Once completed, press the Green Arrow at the bottom right-hand
corner of the screen
(4)
. You will then see the screen below.
In the Value Field, create a name for your test
(1)
. Press the check
mark button
(2)
to save your test name. You will hear an extended
beep sound indicating your test name has been saved. You will see
the following screen appear:
1
2
Before initiating the test sequence, make sure the Ball Seat Adpater
at the bottom of the load cell is positioned no further away from the
Consolidation Cell’s Load Pad than 1/16”. This ensures that the test
will run correctly. If the Ball Seat Adpater is not positioned correctly,
you chance that the distance to it is further than the travel distance of
the machine. See photo on next page.