40
On Tab 3 you will be able to set the number of points you want to
use for your calibration. You can choose 1-point, 5-point, 10-point or
Custom, which allows you to use any number of points up to a total
of 10. In the example on the previous screen shot above, a 5-point
calibration has been chosen.
With your instrumentation sensor placed in a calibration frame and
the sensor plugged into the an Input on the HM-5470. You will set
the “0” point at 0 with no load applied to the sensor. The “5” point
will be set with the maximum load capacity of the sensor applied.
Points “1” through “4” are usually determined by spacing them out
evenly between the no load reading and the maximum load reading.
As an example, if you have a sensor with a 1,000 lb maximum force
capability, you would set the “0” point at “0” and the “5” point at
1,000. Points “1” through “4” would typically be set at: “1” 200; “2”
400; “3” 600; “4” 800 and “5” 1000. Or, divide the maximum load
number by the number of points, in this case 5, which works out to
200 point increments between points. See below.
If you choose Multi-Point (Recommended calibration method) Curve
and click on the Right Arrow in the bottom right-hand corner of the
screen, you will be taken to Tab 3.