It is highly recommended that you use some type of refrigerant or
desiccant dryer and regulator in your machine setup. Humboldt
offers the following items for this purpose.
Refrigeration Dryer, 115V 60Hz 1ph
HM-4221
Compressed air quality is often overlooked in many labs.
Compressed air contains condensate which, when cooled,
will turn into water, causing extensive damage to both the
compressed air network and testing equipment. Refrigera-
tion dryers actively remove this condensate to achieve near
perfectly dry compressed air. The benefits are notable:
less system downtime, reduced costs and maintenance,
and improved test equipment life. This refrigeration dryer,
thanks to its PlusPack heat exchanger and the most com-
pact dimensions on the market, will prove a major asset in
your lab. Dryer uses a 1/2” NPT-F pipe size and nominal
flow is: 10 SCFM, 17 Nm3/hr and 0.3 Nm3/min. based on
an ambient and inlet temperature of 100°F (38°C) and a
working pressure of 100 psig (7 bar)
Desiccant Dryer
HM-4222
Ideal for drying small volumes of air at the point of use.
Convenient in-line mounting saves space. ISO Class
2 dryer. Max. operating pressure is 150 psig. and max
operating temperature is 125°F. Total capacity is 4400 ft
3
,
Female NPT inlet/outlet size is 0.25 NPTF, bowl size is 1.75
lbs, Height 11 Inches, Width 4.625”, Includes one charge
of desiccant
Filter/Regulator HM-4223
One-piece, Filter/Regulator, 0-125 psi (0 - 8.6 bar) with
standard filtration of 5 micron. Height is 9.77 and width
is 2.36. Bowl material is polycarbonate and includes sight
glass and pressure gauge.
HM-4221
HM-4222
HM-4223
Air Connection
The inlet air pressure should not exceed 150psi (1000 kpa). Minimum air
pressure should be set at least 20psi higher than expected operating pres-
sure. For a maximum vertical operating load of 10kN (2000 lb), the pressure
required would be 95psi.