Interfaces and software update
OPERATION MANUAL
Chapter 6
Unichiller®
V2.3.0en/07.06.19//17.12
68
Pin
Signal
1
I+
2
U+
3
U−
4
I−
6.2.3
Service interface
This interface is exclusively used by Huber service engineers for service purposes.
6.3
Interfaces at the Com.G@te® (optional)
Connecting to the interfaces at the temperature control unit during operation
DAMAGE TO THE INTERFACES
When devices in operation are connected with interfaces of the temperature control unit,
interfaces may get damaged.
Before connecting, ensure the temperature control unit and the device to be connected are
turned off.
If the
>Com.G@te<
[46] is not connected, proceed as follows.
PROCEDURE
Switch off the temperature control unit.
Plug the
>Com.G@te<
[46] into the slot provided.
Switch on the temperature control unit. “Unistat Control ONE” detects the new component. The
>Com.G@te<
[46] is operable.
Note that you may only change
>Com.G@te<
[46] when the temperature control unit is turned off.
In the category “Interfaces” you can change the settings of each function such as PoKo, analog
interface and RS232/RS485.
You may order
>Com.G@te<
[46] with Order No. 6915. You can extend an existing
>Com.G@te<
[46] via a connecting cable (Order No. 16160).
Pin assignment (front
view)
Pin assignment
Com.G@te external
interfaces:
LEVEL, POKO, AIF,
ECS,RS232/RS485
(left to right).
Com.G@te internal
interfaces:
POKO, AIF, ECS,
RS232/RS485
(left to right).