Setup mode
OPERATION MANUAL
Chapter 4
Unichiller®
V2.3.0en/07.06.19//17.12
46
4
Setup mode
4.1
Setup mode
Moving the temperature control unit during operation
SERIOUS BURNS/FREEZING OF THE HOUSING PARTS/ESCAPING THERMAL FLUID
Do not move temperature control units that are in operation.
4.1.1
Emergency stop switch (optional): Activate / Deactivate
By activating the
>Emergency stop switch<
[70], all poles of the temperature control unit are
disconnected
immediately
.
4.1.1.1
Emergency stop switch (optional): Activate
PROCEDURE
Press the
>Emergency stop switch<
[70].
All poles of the temperature control unit are disconnected
immediately
.
4.1.1.2
Emergency stop switch (optional): Deactivate
PROCEDURE
Turn the
>Mains isolator<
[36] to the “0” position!
Release the
>Emergency stop switch<
[70] by turning it clockwise. A built-in spring in the
>Emer-
gency stop switch<
[70] returns it to its original state.
4.1.2
Turning on the temperature control unit
PROCEDURE
Switch on the temperature control unit using the
>Mains isolator<
[36].
Only valid for temperature control units with an emergency stop switch (optional):
The safety relay is tested. Press the illuminated
>Release button<
[0]. The temperature control
unit is started and the illumination in the
>Release button<
[0] is turned off.
Then a system test is conducted to test the full functionality of the temperature control unit. All
sensors, the all-pole disconnecting power disconnect relay for the power supply and the power
electronics of the main heater and the main heater itself are checked. A message appears on the
>Touchscreen<
[88] in case of an error or if a warning is issued. Contact Customer Support if in
doubt (the phone number can be found on page 84 in Section
).
Exemplary layout of
the switches
Содержание Unichiller 400T
Страница 1: ...Unichiller...
Страница 2: ......
Страница 3: ...OPERATION MANUAL Unichiller...
Страница 4: ......
Страница 6: ...OPERATION MANUAL Unichiller V2 3 0en 07 06 19 17 12 6 Pilot ONE Layout of the Home screen...
Страница 85: ...Annex Chapter 9 OPERATION MANUAL V2 3 0en 07 06 19 17 12 Unichiller 85 9 Annex...