Setup mode
OPERATION MANUAL
Chapter 4
Unichiller®
V2.3.0en/07.06.19//17.12
60
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Protection Options”.
Tap on the category “Freeze Protection”.
Either tap on the dialog entry “off” or “on” to turn the freeze protection on or off.
Confirm your choice by tapping on “OK”.
4.2.13
Setting the setpoint thresholds
The limits for the minimum and maximum setpoints serve for the safety of your system. They should
be set for the application field of the thermal fluid before starting the first temperature control and
when changing the thermal fluid. The maximum setpoint limit limits the setpoint setting for the bath
or flow temperature. The minimum setpoint limit protects against high viscosity or freezing at low
temperatures. Setpoint settings are only available in the temperature range between the minimum
and maximum setpoint limit.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Protection Options”.
Tap on the category “Setpoint Limits”.
Tap on the sub-category “Minimum Setpoint”.
Enter the new value, using the numeric keypad that appears.
Confirm your entry by tapping on “OK”.
In the display that follows, confirm your entry again by tapping on “OK”.
The correct selection will be displayed graphically and the “Min. setpoint” will be changed
promptly. If tapping on “OK” is not correct, it will be displayed graphically for 2 seconds. The dis-
play will then return to the category “Setpoint Limits”. Try changing the “Min. Setpoint” again.
Tap on the sub-category “Maximum Setpoint”.
Enter the new value, using the numeric keypad that appears.
Confirm your entry by tapping on “OK”.
In the display that follows, confirm your entry again by tapping on “OK”.
The correct selection will be displayed graphically and the “Max. Setpoint” will be changed
promptly. If tapping on “OK” is not correct, it will be displayed graphically for 2 seconds. The dis-
play will then return to the category “Setpoint Limits”. Try changing the “Max. Setpoint” again.
Check the set values for the minimum and maximum setpoint at any system change, especially
when changing the thermal fluid.
4.2.14
Setting the setpoint
PROCEDURE
Go to the “Home” screen.
Tap on the keypad symbol next to “T
Setpoint
”.
Overview of the
temperature thresh-
olds
Содержание Unichiller 400T
Страница 1: ...Unichiller...
Страница 2: ......
Страница 3: ...OPERATION MANUAL Unichiller...
Страница 4: ......
Страница 6: ...OPERATION MANUAL Unichiller V2 3 0en 07 06 19 17 12 6 Pilot ONE Layout of the Home screen...
Страница 85: ...Annex Chapter 9 OPERATION MANUAL V2 3 0en 07 06 19 17 12 Unichiller 85 9 Annex...