FULLTEST3
EN - 93
FUNCTIONAL TEST – TRMS LOAD CURRENT ON SOCKET
Range [A]
Resolution [A]
Accuracy
Overload
protection
0.00
y
0.99
0.01
± (3%rdg. + 3dgt)
CAT II 300 V
1.0
y
20.0
0.1
Frequency range:
15Hz
y
723Hz
Protection:
automatic interruption 10s after exceeding 16A
Pre-Test
Grounding of PE terminal on schuko socket
FUNCTIONAL TEST – TRMS MAINS VOLTAGE ON SOCKET
Range [V]
Resolution [V]
Accuracy
Overload
protection
195
y
253
1
± (2%rdg + 2dgt)
CAT II 300 V
Frequency range
15Hz
y
723Hz
FUNCTIONAL TEST – APPARENT POWER / ACTIVE POWER ON SOCKET
Range [VA/W]
Resolution
[VA/W]
Accuracy
Overload
protection
0.0
y
99.9
0.1
±(5%rdg. + 10dgt)
CAT II 300 V
100
y
999
1
± (5%rdg. + 3dgt)
1.00
y
5.06 k
10
Limit value PAPP
Adjustable 6
y
999VA, 1.00kVA
y
5.06kVA
FUNCTIONAL TEST – POWER FACTOR ON SOCKET
Range
Resolution
Accuracy
Overload
protection
0.00
y
1.00
0.01
Depend on P
APP
and P
ACT
CAT II 300 V
FUNCTIONAL TEST – TRMS LEAKAGE CURRENT ON SOCKET
Range
Resolution
Accuracy
Overload
protection
0.25
y
19.99mA
0.01mA
± (3%rdg. + 3dgt)
CAT II 300 V
20.0
y
49.9mA
0.1mA
0.05
y
0.99A
0.01A
1.0
y
10.0A
0.1A
Load current influence
± 0.01 mA/A
Limit value:
0.25mA
y
10.00A (standard limit value 3.50mA)
Frequency range
40Hz
y
723Hz (characteristics according to IEC 61557-13)
Measurement method
Differential current
Mains cord polarity exchange
yes
UUT mains on switch test
yes (limit value 25mA)
Over-range limit (IPE)
yes (measurement interrupted after 10s after exceeding 10A)
Over-range limit (IL)
yes (measurement interrupted after 10s after exceeding 16A
PHASE SEQUENCE (PHASESEQ)
Range [V]
Resolution [V]
Accuracy
Overload
protection
360
y
460
1
±(2%rdg + 2dgt)
CAT III 300 V
Display of test result:
1.2.3 (correct) or 2.1.3 (incorrect) or 1.1.X (not defined)