FULLTEST3
EN - 73
6.12. LEAKAGE CURRENT (ILEAK)
The instrument allows the measurement of the AC leakage current both with the use
of a transducer clamp (optional accessory HT96U) connected to the ILEAK input (see
Fig. 2 - part 26) and on equipment directly connected to the schuko test socket on the
front panel (see Fig. 2 - part 25). In this case the instrument powers the UUT and
measures the leakage current on the plug.
1. Press the
FUNC
key and select the
ILEAK
function. The following screen is
shown on the display
Fig. 71 : ILEAK initial screen
2. Select the test parameters on the instrument (see Table 13) and carry out the
desired setup
Parameter Description
Value
MODE
Measurement mode
CLAMP or SOCKET
RANGE
Measurement range of clamp
1000mA,100A,1000A
LIMIT
Limit value on measurement
with HT96U
0.1mA
y
1000mA (1000mA)
0.1A
y
100.0A (100A)
1.0A
y
1000A (1000A)
Limit value on measurement
with connection to test socket
0.25mA ÷ 10.0A
L POS
Position of Phase terminal on the
schuko test socket
LEFT / RIGHT
Table 13 : Setup parameters of ILEAK
function
Use of transducer clamp HT96U
3. Select CLAMP mode by pressing the
MODE
touch-screen key first
4. Check selected measurement range and modify it if needed by pressing the
RANGE
touch-screen key first
5. Check selected limit leakage current by pressing the
LIMIT
touch-screen key first.
Four independent limit values are available for quicker operations. Select the
closest one and modify it by using the
+
and
ʊ
touch screen keys if needed
6. Select measurement screen by pressing the
touch-screen key and check all
settings again
7. Connect the current clamp to measured wiring according to the