FULLTEST3
EN - 49
CAUTION
x
When selecting RCD type (TYPE), nominal differential current
(I
'
N
) or measurement (MEAS) it may happen that wished
parameter will not be available (see § 10.5). In this case the level
of other parameter or even other two parameters shall be
reduced first
x
In case both voltages U
L/N
and U
L/PE
within required range 100
y
265V are present at L/N/PE test terminals (also displayed) but
there is no READY message displayed, check if energizing
socket is correctly earthed
6.5.1. Anomalous situations
The following specific information can be shown on the display during measurement
Information displayed
Description
VOLTAGE OUT OF RANGE
Input voltage U
L/N
or U
L/PE
out of required
range 100
y
265 V after pressing the START
key.
MEASUREMENT FAILED!
Input voltage failed during the measurement
(disconnection of test leads, installation fuse
tripped etc.)
CONTACT VOLTAGE!
Contact voltage higher than set limit value (25
V or 50 V)
EXTERNAL IMPEDANCE
TOO HIGH!
Impedance in L conductor is too high, preset
current can not be generated.
FUSE F3!
Fuse F3 is blown.
HOT!
Internal circuitry is overheated.
Wait to cool it down!