FULLTEST3
EN - 44
Item
Description
9
Leakage current in green colour if the result is lower than or equal to
set limit value. If break-through occurred during the test then limit
value will be displayed in red colour
10
Real time clock (hh.mm.ss).
11
Unit of measurement result (mA)
12
Set measurement time (in RAMP mode only)
13
Measurement result status (symbol
in green colour - result lower
than or equal to set limit value, symbol
in red colour - break
through occurred during the test or result higher than set limit value)
10. A warning with explanation of how to connect test leads with regard to selected
test voltage will be displayed. Check the connection then confirm it by pressing
the
YES
touch-screen key, the message “READY” will appear and will stay
present for 10 seconds.
START/STOP
button is active while “READY” message is
present. Press and keep pressing the
START/STOP
button, test voltage will be
applied to test terminals. The test will be stopped after releasing the
START/STOP
button (MANUAL or BURN mode) or after elapsing set test time
(RAMP mode). In PULSE mode, press and keep the
START/STOP
button for
at
least 5s
until the result is shown on the display
11. Test result will currently be displayed in green color if it is lower than or equal to
set limit value. Final result will be equipped with green
symbol and with beep-
beep sound (result OK). If break through occurred during the test then the test will
be stopped and limit test current will be displayed in red color equipped with red
symbol and with longer beep sound
12. Save the test result by pressing the
SAVE
key (see § 7.1)
Содержание FULLTEST3
Страница 1: ... Copyright HT ITALIA 2020 Release EN 2 01 13 01 2020 FULLTEST3 User manual ...
Страница 2: ......
Страница 99: ......