Tighten the pressurisation block as shown in
the figure.
10.
Variation ES939
: reassemble the pneumatic
block, tightening the four screws indicated in
the figure.
11.
Apply the two pneumatic connectors.
12.
Close the internal movable connections
(inductive sensors and electric fan) and insert
the rubber pressurisation tubes in the special
connector on the pneumatic connections block
13.
Carry out the sensor calibration procedure.