Digitizer Command Reference
93
Gain DAC Test (self-test numbers 63-66)
(HP E1564A 4-Channel Digitizer)
Uncalibrated Gain Test (self-test numbers 67-94)
(HP E1564A 4-Channel Digitizer)
:TST[:RESults]?
TEST:TST[:RESults]? returns an array of integers that result from the self-test
command *TST?.
Comments
•
A response of “0” indicates there is no error.
•
Use the TEST:ERR? <test_number> command to retrieve details about the
failed test number(s) returned by the TEST:TST:RESults? command.
BCD weight
Failure mode
1
DAC measurement is noisy
2
Measured data span is too small
4
Lower end point to upper end point span is too small
8
Lower end point to upper end point span is too large
16
Gain DAC span does not include 0
32
Bit weight is out of limits; the offending bit is in B15-B8.
64
Gain DAC nominal setting is out of limits.
BCD weight
Failure mode
1
The max-to-min span is 0.0.
2
Gain span is too large.
4
Gain mean is too low.
8
Gain mean is too high.
16
Gain standard deviation is too large.
32
Gain is out of limits.
Содержание E1563A
Страница 8: ......
Страница 24: ...24 Digitizer Module Set up ...
Страница 84: ...84 Digitizer Command Reference ...
Страница 110: ...110 ...
Страница 138: ...138 HP E1563A and E1564A Register Based Programming Appendix B ...
Страница 156: ...156 HP E1563A and E1564A Verification Tests ...