Digitizer Command Reference
75
ROSCillator:SOURce
[SENSe:]ROSCillator:SOURce INTernal | EXTernal is used to specify the
timebase source. The default timebase is the INTernal timebase which uses the VXI
CLK10, 10 MHz reference. The EXTernal input is the TTL “Time Base” input pin
on the front panel D-subminiature connector (right pin column, bottom pin).
HP E1563A
HP E1564A
(“Time Base” input - bottom right pin)
Note
The EXTernal source requires you also send the
ROSC:EXT:FREQ <
freq>
command to specify the frequency of the external timebase.
Comments
•
The timebase reference set by
SAMPle:TIMer <
interval
>
is used when the
sample source is TIMer (
SAMPle:SOURce TIMer
).
•
Executable when initiated: NO
•
Coupled command: Yes, The
SAMPle:TIMer <
interval
>
is set to a period or
interval nearest the old value when source is changed from EXTernal to
INTernal or vice versa.
•
Reset (*RST) Condition: INTernal source, freq = 10.0 MHz
ROSCillator:SOURce?
[SENSe:]ROSCillator:SOURce? queries to determine the timebase source.
Returns either INTernal or EXTernal.
SWEep:POINts
[SENSe:]SWEep:POINts
<count> | MIN | MAX command sets the number of
sweep points. The number of points set is common to all channels. You cannot have
two different channels with different a sweep point count.