
42
Digitizer Command Reference
ABORt
This command aborts a measurement in progress or stops a measurement being made
continuously. The command is ignored without error if a measurement is not in
progress. This command also aborts a calibration in progress and will set the
CAL:STATe to OFF.
Subsystem Syntax
ABORt
Comments
•
Use the DATA:COUNt? query to determine how many readings were taken
before the ABORt was received.
•
ABORt does not affect any instrument settings.
•
Executable when initiated: YES
•
Coupled command: No
•
Reset (*RST) Condition: None
Содержание E1563A
Страница 8: ......
Страница 24: ...24 Digitizer Module Set up ...
Страница 84: ...84 Digitizer Command Reference ...
Страница 110: ...110 ...
Страница 138: ...138 HP E1563A and E1564A Register Based Programming Appendix B ...
Страница 156: ...156 HP E1563A and E1564A Verification Tests ...