Digitizer Application Information
33
Figure 2-5. Slave Module Configuration Block Diagram.
TRIG:MODE
SLAVe0 pairs TTLT0 (sample) with TTLT1 (trigger)
A SLAVe0 module will function with other SLAVe0 modules and
the MASTer0 module.
1) The trigger source from the slave can be set
with
TRIG:SOURce2 IMM | INT1-4 | EXT |
TTLT<n>
.
2) SLAVe0 sets the TTLT0 line as if it were
SAMP:SOUR TTLT0
and sets the TTLT1 line as
if it were
TRIG:SOUR1 TTLT1
. However, the
queries
SAMP:SOUR?
or TRIG:SOUR? will not
return these settings. These lines are simply
dedicated for synchronization between the
modules in the master- slave mode. You should
not use these lines for any other purpose with the
OUTPut, SAMPle or TRIGger commands.
MODE
SLAVe
Sample signal
SLAVe0
TTLT0
SLAVe2
TTLT2
SLAVe4
TTLT4
SLAVe6
TTLT6