Digitizer Command Reference
45
•
Executable when initiated: No
•
Coupled Command: YES; Range changes will change the value; the
percent of full scale of the range will be kept constant.
For example, on the 4 volt range, with a 2V limit, a
range change to 16V will set a new limit of 8V.
•
Related Command: [SENSe:]VOLTage[<channel>][:DC]:RANGe <range>
•
Reset (*RST) Condition: -254 volts
:LIMit:LOWer:DATA?
CALCulate[<channel>]:LIMit:LOWer:DATA? [MIN | MAX] queries the
lower limit value set for the specified channel.
:LIMit:UPPer[:STATe]
CALCulate[<channel>]:LIMit:UPPer[:STATe] OFF | 0 | ON | 1 enables the
upper limit checking for the specified channel. Use :LIMit:UPPer:DATA <value>
to set the actual limit value to be tested against.
Comments
•
Executable when initiated: YES
•
Coupled command: YES, setting the upper state ON will cause
LIMit:LOWer[:STATe] to be set OFF (if it is ON).
Note
An error will be generated if you have TRIG:SOURce set to INT1-4 and the
internal input is the same as the channel you are attempting to enable the upper limit
testing. For example, if TRIG:SOUR INT2 is set. The trigger level from channel 2
is the trigger event that is the internal trigger input.
CALC:LIMit:UPPer:STATe ON is attempting to use this signal for limit testing
and creates a settings conflict. Either the trigger level can be used as an internal
trigger or the level can be used in limit testing, but not both.
•
Reset (*RST) Condition: OFF
:LIMit:UPPer[:STATe]?
CALCulate[<channel>]:LIMit:UPPer[:STATe]? queries the upper limit
checking state to see if it is enabled or disabled for the specified channel. This
command returns the voltage level measured and the detection mode. A returned
value of “0” indicates the specified channel is disabled for upper limit checking. “1”
returned indicates the specified channel is enabled and will detect signals above the
specified upper limit.