17. Signal ID Oscillator Adjustment
9. While observing the HP
display, adjust
SIG ID for the highest
obtainable frequency with less than 3
decrease in amplitude from maximum. Read this
frequency from the frequency counter and record as
HIGH =
Hz
10. Observe the HP
display as you adjust
SIG ID for the lowest obtainable
frequency with less than 3
decrease in amplitude from maximum. Record the
frequency counter reading as
L O W
=
MHz
11. Calculate the difference between
and
then divide results by four.
Enter the result as
=
12. Add
to
recorded in step 10 and record the result as
F
=
MHz
13. Adjust
for a frequency counter reading equaling
The final adjusted
frequency must equal 10.7 MHz
Adjustment Procedures 2-65
Содержание 8560A
Страница 93: ......
Страница 150: ... ...
Страница 178: ...4 20 Replaceable Parts ...
Страница 180: ...4 22 Replaceable Parts ...
Страница 182: ...4 24 Replaceable Parts ...
Страница 184: ...4 26 Replaceable Parts ...
Страница 186: ...4 28 Replaceable Parts ...
Страница 188: ...4 30 Replaceable Parts ...
Страница 190: ...4 32 Replaceable Parts ...
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 194: ......
Страница 195: ......
Страница 200: ...w 5 4 A 4 w 5 5 W 2 7 Figure 5 3 Top View A2 and A3 Unfolded SK157 Major Assembly and Cable Locations 5 5 ...
Страница 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Страница 249: ...6 44 General Troubleshooting ...
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 282: ......
Страница 307: ......
Страница 319: ......
Страница 324: ...8 42 IF Section ...
Страница 325: ......
Страница 326: ......
Страница 327: ......
Страница 344: ......
Страница 397: ......
Страница 398: ......
Страница 423: ......
Страница 441: ...12 18 Display Power Supply Section ...
Страница 442: ......
Страница 443: ......
Страница 449: ...A 6 Component Level Information Packets ...
Страница 450: ......