4. IF Amplitude Adjustments
Procedure
1. Set the HP 8560A
switch to off. Remove the analyzer cover and place the analyzer
in the service position as illustrated in Figure 2-6. See Figure 2-7 for adjustment location.
2. Disconnect W29, violet coax cable, from
Connect the test cable between
and
the 50 output of the HP 3335A. Set the HP 8560A (LINE) switch on.
3. Set the HP 8560A controls as follows:
C E N T E R F R E Q
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200kHz
REF LVL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..ldB/DIV
RES BW
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300kHz
VIDEO BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4. On the HP
press
and IF
ADJ OFF.
5. Set the HP 3335A controls as follows:
FREQUENCY
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -55dBm
6. Note the marker value. Ideally it should read -60
+/-.l
7. If the marker reads below -60.1
rotate
CAL OSC AMPTD one-third turn
clockwise for every 0.1
below -60
See Figure 2-7 for the location of
8. If the marker reads above -59.9
rotate
CAL OSC AMPTD one-third turn
counter clockwise for every 0.1
above -60
A change in the displayed amplitude
will not be seen at this point.
Note
If
has inadequate range, refer to “Inadequate CAL OSC AMPTD
Range” in Chapter 9.
9. Press
CURR IF STATE
. After allowing the analyzer time to complete the
adjustments, the displayed amplitude and marker reading should change.
10. Repeat steps 7 and 8 until the marker reads -60
+/-.l
11. Disconnect the test cable from
and reconnect W29 to
Adjustment Procedures 2-27
Содержание 8560A
Страница 93: ......
Страница 150: ... ...
Страница 178: ...4 20 Replaceable Parts ...
Страница 180: ...4 22 Replaceable Parts ...
Страница 182: ...4 24 Replaceable Parts ...
Страница 184: ...4 26 Replaceable Parts ...
Страница 186: ...4 28 Replaceable Parts ...
Страница 188: ...4 30 Replaceable Parts ...
Страница 190: ...4 32 Replaceable Parts ...
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 194: ......
Страница 195: ......
Страница 200: ...w 5 4 A 4 w 5 5 W 2 7 Figure 5 3 Top View A2 and A3 Unfolded SK157 Major Assembly and Cable Locations 5 5 ...
Страница 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Страница 249: ...6 44 General Troubleshooting ...
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 282: ......
Страница 307: ......
Страница 319: ......
Страница 324: ...8 42 IF Section ...
Страница 325: ......
Страница 326: ......
Страница 327: ......
Страница 344: ......
Страница 397: ......
Страница 398: ......
Страница 423: ......
Страница 441: ...12 18 Display Power Supply Section ...
Страница 442: ......
Страница 443: ......
Страница 449: ...A 6 Component Level Information Packets ...
Страница 450: ......