10. If log fidelity is poor near the bottom of the screen or the 1 MHz resolution bandwidth is
narrow, a fault might exist in the wide/narrow filter switch. Refer to function block G
of A4 Log Amplifier Schematic Diagram in the
HP
Spectrum Analyzer
Component Level Information
binder. Check this switch as follows:
a. Monitor voltages on
pins 1 and 7 while changing the spectrum analyzer
resolution bandwidth from 100
to 300
b. If the voltages do not come within a few volts of the
V and -15 V supplies,
and
are suspect.
c. Disconnect the digital multimeter and reconnect W27 to
Linear Amplifiers
Refer to function block C of A4 Log Amplifier Schematic Diagram (sheet 2 of 4) in the
HP
Spectrum Analyzer Component Level Information
binder.
The Linear Amps consist of two variable gain stages,
and
as well as the buffer
amplifier
and
The linear amplifiers provide 0 to 40
of IF
gain in 10
steps. The gain of
can be increased by 20
by turning on
and
with the control line LIN_20B. The gain of
can be increased by either
10
or 20
with the control lines LIN_10 or LIN_20A respectively. The gain can be
selected by setting the spectrum analyzers reference level.
IF Gain Application Guidelines
Power into
Reference Level Gain of
Gain of
Total Gain
- 5 0
0
0
- 4
- 6 0
0
10
10
14dBm - 7 0
0
20
20
-24
- 8 0
20
10
30
-34
- 9 0
20
20
40
Total gain can be measured by injecting the specified power into
and measuring the
total gain provided by
and
The following procedure provides a means of
troubleshooting the linear amplifiers.
1. On the spectrum analyzer press (PRESET), (SPAN), ZERO SPAN ,
1
50
LINEAR,
MORE, AMPTD UNITS
,
and
2. Disconnect W27 (coax 3) from
and connect W27 to the output of a signal generator.
3. Set the signal generator controls as follows:
Amplitude
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4. Simultaneously decrease the signal generator output and spectrum analyzer REF LVL in
10
steps to -90
At each step, the signal displayed on the spectrum analyzer
should be within one division of the previous position.
5. If a problem exists, isolate it by comparing the actual gain of
and
with
those listed in the above gain guidelines.
6. Reconnect W27 (coax 3) to
IF Section 8-15
Содержание 8560A
Страница 93: ......
Страница 150: ... ...
Страница 178: ...4 20 Replaceable Parts ...
Страница 180: ...4 22 Replaceable Parts ...
Страница 182: ...4 24 Replaceable Parts ...
Страница 184: ...4 26 Replaceable Parts ...
Страница 186: ...4 28 Replaceable Parts ...
Страница 188: ...4 30 Replaceable Parts ...
Страница 190: ...4 32 Replaceable Parts ...
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 194: ......
Страница 195: ......
Страница 200: ...w 5 4 A 4 w 5 5 W 2 7 Figure 5 3 Top View A2 and A3 Unfolded SK157 Major Assembly and Cable Locations 5 5 ...
Страница 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Страница 249: ...6 44 General Troubleshooting ...
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 282: ......
Страница 307: ......
Страница 319: ......
Страница 324: ...8 42 IF Section ...
Страница 325: ......
Страница 326: ......
Страница 327: ......
Страница 344: ......
Страница 397: ......
Страница 398: ......
Страница 423: ......
Страница 441: ...12 18 Display Power Supply Section ...
Страница 442: ......
Страница 443: ......
Страница 449: ...A 6 Component Level Information Packets ...
Страница 450: ......