307
ACC
308 FREQ ACC
309
ACC
310
ACC
311
ACC
312
ACC
313 FREQ ACC
314 FREQ ACC
Transfer oscillator pretuned DAC out of range. The transfer oscillator
pretune procedure attempts to find pretuned DAC values by
programming the PLL to 25 different frequencies and incrementing
the transfer oscillator pretune DAC until XFRSENSE changes
polarity. ERR 307 is set if the DAC is set to 255 (maximum) before
XFRSENSE changes polarity.
Offset oscillator pretune DAC not within prescribed limits at low
frequency. The offset oscillator pretune DAC is set to provide a
frequency less than 189 MHz while the PLL is programmed for 189
MHz. ERR 308 is set if XFRSENSE is greater than
V (it should
be at the negative rail).
Offset oscillator pretune DAC not within prescribed limits at high
frequency. The offset oscillator pretune DAC is set to provide a
frequency less than
MHz while the PLL is programmed for 204
MHz. ERR 309 is set if XFRSENSE is greater than
V (it should
be at the negative rail).
Main roller pretune DAC set to 255. The main roller pretune DAC is
set to 5, causing MAINSENSE to go to the positive rail. The DAC
is incremented until MAINSENSE changes polarity. ERR 310 is set
if the DAC is set to 255 before MAINSENSE changes to a negative
polarity.
Main roller pretune DAC set to 255. The main roller pretune DAC is
set to 5, causing MAINSENSE to go to the positive rail. The DAC
is incremented until MAINSENSE changes polarity. ERR 311 is set
if the DAC is set to 255 before MAINSENSE changes to a negative
polarity.
Unable to adjust MAINSENSE to 0 volts using the fine adjust DAC.
The coarse adjust and fine adjust DAC are used together to set
MAINSENSE to 0 volts with the loop opened. ERR 312 is set if the
fine adjust DAC cannot bring MAINSENSE to 0 volts.
Error in LO synthesis algorithm. ERR 313 is set if a combination of
Sampler oscillator and roller oscillator frequencies could not be found
to correspond to the required YTO start frequency. Contact the
factory.
Indicates problems in the span calibration. Troubleshoot any unlocks
before attempting to troubleshoot span calibration problems, because
the loops must all lock in order to perform the calibration. If LO
spans greater than 1 MHz are correct, check
,
or
This error message appears when the main
roller oscillator sweep sensitivity is 0. A sweep ramp is injected into
the locked main roller loop which should generate a negative-going
ramp on MAINSENSE. ERR 314 is set if the slope of this ramp is 0.
This is an indication of an unlocked main roller loop or lack of a sweep
ramp.
General Troubleshooting 6-19
Содержание 8560A
Страница 93: ......
Страница 150: ... ...
Страница 178: ...4 20 Replaceable Parts ...
Страница 180: ...4 22 Replaceable Parts ...
Страница 182: ...4 24 Replaceable Parts ...
Страница 184: ...4 26 Replaceable Parts ...
Страница 186: ...4 28 Replaceable Parts ...
Страница 188: ...4 30 Replaceable Parts ...
Страница 190: ...4 32 Replaceable Parts ...
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 194: ......
Страница 195: ......
Страница 200: ...w 5 4 A 4 w 5 5 W 2 7 Figure 5 3 Top View A2 and A3 Unfolded SK157 Major Assembly and Cable Locations 5 5 ...
Страница 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Страница 249: ...6 44 General Troubleshooting ...
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 282: ......
Страница 307: ......
Страница 319: ......
Страница 324: ...8 42 IF Section ...
Страница 325: ......
Страница 326: ......
Страница 327: ......
Страница 344: ......
Страница 397: ......
Страница 398: ......
Страница 423: ......
Страница 441: ...12 18 Display Power Supply Section ...
Страница 442: ......
Страница 443: ......
Страница 449: ...A 6 Component Level Information Packets ...
Страница 450: ......