Procedure 7.
High Voltage Assembly
Removal
Warning
The A6 Power Supply and
High Voltage assemblies contain lethal voltages
with lethal currents in all areas. Use extreme care when servicing these
assemblies. Always disconnect the power cord from the instrument before
beginning this replacement procedure. Failure to follow this precaution can
represent a shock hazard which may result in personal injury.
1. Disconnect the power cord from the analyzer.
2. Remove the analyzer’s cover assembly as described in Procedure 1.
3. Fold out the A2, A3, A4, and A5 assemblies as described in Procedure 5,
A3, A4,
and A5 Assemblies.”
4. Place the analyzer top-side-up on the work bench.
Warning
The voltage potential at
is
Disconnect at the CRT with caution!
Failure to properly discharge
may result in
severe electrical shock
to
personnel and damage to the instrument.
5. Connect the analyzer’s line-power cord to provide proper grounding while discharging the
post-accelerator cable. Make sure that the analyzer’s line-power switch is in the
off position.
6. Connect a high-voltage probe
such as the HP 34111A to a voltmeter with a
10
input.
7. Connect the clip lead of the probe (ground) to the chassis of the spectrum analyzer.
8. Slip the tip of the high-voltage probe under the
post-accelarator cable’s rubber
shround to obtain a reading on the voltmeter. See Figure 3-2.
9. Keep the high-voltage probe on the post-accelerator connector until the voltage has
dropped to a voltmeter reading of less than 5
(less than 5 V at the connector). This
normally takes about 30 seconds.
10. Disconnect the line-power cord from the spectrum analyzer.
11. Using a small screwdriver with the shank in contact with the CRT shield assembly, slip
the tip of the screwdriver under the
post-accelerator cable’s rubber shroud and
short the cable to ground on the CRT shield assembly. See Figure 3-2.
12. Pry out the black grommet protecting post-accelerator cable
from the CRT
shield assembly.
13. Carefully unsnap the
post-accelerator cable from the CRT and discharge it by
shorting the cable to chassis ground on the CRT shield assembly.
14. Remove the three screws securing the power supply shield to the power supply and remove
the shield.
15. Remove the three screws and washers securing the
High Voltage assembly to the A6
Power Supply assembly.
3-22 Assembly Replacement
Содержание 8560A
Страница 93: ......
Страница 150: ... ...
Страница 178: ...4 20 Replaceable Parts ...
Страница 180: ...4 22 Replaceable Parts ...
Страница 182: ...4 24 Replaceable Parts ...
Страница 184: ...4 26 Replaceable Parts ...
Страница 186: ...4 28 Replaceable Parts ...
Страница 188: ...4 30 Replaceable Parts ...
Страница 190: ...4 32 Replaceable Parts ...
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 194: ......
Страница 195: ......
Страница 200: ...w 5 4 A 4 w 5 5 W 2 7 Figure 5 3 Top View A2 and A3 Unfolded SK157 Major Assembly and Cable Locations 5 5 ...
Страница 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Страница 249: ...6 44 General Troubleshooting ...
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 282: ......
Страница 307: ......
Страница 319: ......
Страница 324: ...8 42 IF Section ...
Страница 325: ......
Страница 326: ......
Страница 327: ......
Страница 344: ......
Страница 397: ......
Страница 398: ......
Страница 423: ......
Страница 441: ...12 18 Display Power Supply Section ...
Страница 442: ......
Страница 443: ......
Страница 449: ...A 6 Component Level Information Packets ...
Страница 450: ......