Flatness Out-of-Tolerance
The output level flatness of the tracking generator is specified at a 0
output power
setting. In general, most flatness problems will be a result of a failure in the A10 Tracking
Generator microcircuit. However the POWER LVL signal from the Al4 Frequency Control
Assembly and the
LO IN signal from the A7 LODA can also contribute to flatness
problems.
1. Check the function of the POWER LVL signal from the Al4 Frequency Control Assembly.
Set the TRK GEN RF POWER to a level at which the flatness is out-of-tolerance.
Monitor
pin 1 with a DVM and step the Center Frequency in 100 MHz increments
from 100 MHz to 2.9
and plot the voltage variation versus frequency. A discontinuity
in the plot near the frequency at which the flatness is out-of-tolerance indicates a problem
on the Al4 Frequency Control Assembly.
2. Check the flatness of the
LO IN signal. Perform the
LO OUTPUT Amplitude
performance test. If the test passes, the fault is most likely in the A10 Tracking
Generator. If the test fails, note the Center Frequency setting at which the power level
was out-of-tolerance and compare against the frequency(ies) at which the flatness was
out-of-tolerance. Repeat the test with the power sensor connected to the end of W43
nearest
noting the Center Frequency of any out-of-tolerance power levels. The power
level should be
If the power level is correct at W43 but out-of-tolerance at the
LO OUTPUT and the
Center Frequency setting of the out-of-tolerance power levels is close to the frequencies at
which the output is unleveled, suspect
If the power level at W43 is also out-of-tolerance, suspect either the A7 LODA or the All
YTO. Refer to Synthesizer Section Troubleshooting.
3. Check all coax cables, especially semi-rigid cables. A fault in one of these cables can cause
a very high-Q power hole.
Vernier Accuracy Out-of-Tolerance
Vernier accuracy is a function of the POWER LVL drive signal from the Al4 Frequency
Control Assembly and the ALC circuitry on
The vernier accuracy is specified at
300 MHz. Since vernier accuracy is tested using a broadband power sensor, abnormally high
spurious responses could cause the measured vernier accuracy to fail when in fact the accuracy
of the 300 MHz signal alone is within specification.
1. Check the POWER LVL drive signal from A14. Monitor
pin 1 with a DVM.
Change the TRK GEN RF POWER in 1
steps and note the voltage at each power
level setting. The voltage should change by the same amount for each 1
step. If the
voltage does not change by the same amount for each 1
step, the fault lies on the Al4
Frequency Control Assembly.
2. Check for abnormally high spurious outputs. Connect the RF OUT
to the input of
another spectrum analyzer (the “test analyzer”). Set the test analyzer to sweep from
300
to 2.9
with a sweeptime of 100
or less. Set the HP 8560A to sweep
from 300
to 2.9
with a 50 second sweeptime. Press
on the HP 8560A
and observe any responses on the test analyzer, ignoring the desired output signal. If any
spurious responses are greater than -20
the vernier accuracy measurement may fail.
Refer to Harmonic/Spurious Outputs Too High.
RF Section 11-21
Содержание 8560A
Страница 93: ......
Страница 150: ... ...
Страница 178: ...4 20 Replaceable Parts ...
Страница 180: ...4 22 Replaceable Parts ...
Страница 182: ...4 24 Replaceable Parts ...
Страница 184: ...4 26 Replaceable Parts ...
Страница 186: ...4 28 Replaceable Parts ...
Страница 188: ...4 30 Replaceable Parts ...
Страница 190: ...4 32 Replaceable Parts ...
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 194: ......
Страница 195: ......
Страница 200: ...w 5 4 A 4 w 5 5 W 2 7 Figure 5 3 Top View A2 and A3 Unfolded SK157 Major Assembly and Cable Locations 5 5 ...
Страница 205: ...Bl Figure 5 8 Rear View BTl A20 5 10 Major Assembly and Cable Locations ...
Страница 249: ...6 44 General Troubleshooting ...
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 282: ......
Страница 307: ......
Страница 319: ......
Страница 324: ...8 42 IF Section ...
Страница 325: ......
Страница 326: ......
Страница 327: ......
Страница 344: ......
Страница 397: ......
Страница 398: ......
Страница 423: ......
Страница 441: ...12 18 Display Power Supply Section ...
Страница 442: ......
Страница 443: ......
Страница 449: ...A 6 Component Level Information Packets ...
Страница 450: ......