2-12
HP 85106D mm-Wave System
Operation
Operating Notes
Important changes caused by using “Factory Preset” to the desired mm-wave
instrument state are as follows:
After factory preset, the operator must recall an appropriate mm-wave
instrument state previously saved. This is usually instrument State 8.
RF Signal Power
Control
The RF signal source power is set to a nominal value by the system
configuration software. If needed, adjust the level for your application. For
maximum measurement accuracy, calibrate the system with the same source
power as to be used during the measurement. RF source power is adjusted by
pressing the keys in the Stimulus [
MENU
], {
POWER MENU POWER SOURCE 1
}.
WARNING
If the network analyzer displays
I.F. OVERLOAD
warning, decrease the RF
source power.
SWEEP MODE
Factory preset selects step sweep mode.
The mm-wave system operates in Step,
Single Point, or Frequency List sweep
modes only.
SET Z
0
Factory preset selects 50; the mm-wave
system uses 1.
WAVEGUIDE DELAY
Factory preset selects the coaxial
configuration. For mm-wave measurements,
select the waveguide configuration.
WAVEGUIDE CUTOFF
This function defines the waveguide cutoff
frequency, f
co
used in the electrical delay
computation. Factory preset selects 0 Hz.
SOURCE #1 POWER and SOURCE #2
POWER
Power levels defined in the standard
instrument state depend on the source and
waveguide frequency band; the levels
should be close to those listed in the
sections titled "RF Signal Power Control"
and "LO Signal Power Control," and Table
on page 2-14.
Содержание 85106D
Страница 11: ...x HP 85106D mm Wave System ...
Страница 95: ...4 2 HP 85106D mm Wave System Service and Troubleshooting Figure 4 1 HP 85106D System Block Diagram ...
Страница 96: ...HP 85106D mm Wave System 4 3 Service and Troubleshooting Figure 4 2 HP 85106D Option 001 System Block Diagram ...
Страница 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Страница 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 149: ...5 36 HP 85106D mm Wave System HP 85104A Test Set Module Assembly Procedure Figure 5 12 HP 85104A Assembly Procedure ...
Страница 153: ...5 40 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 13 HP Q85104A Replaceable Parts ...
Страница 156: ...HP 85106D mm Wave System 5 43 HP 85104A Test Set Module Replaceable Parts Figure 5 14 HP U85104A Replaceable Parts ...
Страница 159: ...5 46 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 15 HP V85104A Replaceable Parts ...
Страница 162: ...HP 85106D mm Wave System 5 49 HP 85104A Test Set Module Replaceable Parts Figure 5 16 HP W85104A Replaceable Parts ...
Страница 208: ......
Страница 212: ...6 48 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 28 HP 85105A Major Assemblies ...
Страница 214: ...6 50 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 29 HP 85105A Semi rigid RF Cables ...
Страница 218: ...6 54 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 31 Wire Harnesses ...
Страница 220: ...6 56 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 32 HP 85105A Front Panel ...
Страница 222: ...6 58 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 33 Rear Panel Standard Option 050 ...
Страница 228: ...6 64 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 37 HP 85105A Accessories Standard ...
Страница 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Страница 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Страница 244: ... vi HP 85106D mm Wave System ...