HP 85106D mm-Wave System
6-33
HP 85105A mm-Wave Controller
LO ALC Adjustment
1. Run the system software and select {
HIGHBAND
} to make the mm-wave
controller active.
2. Check the user parameter (al or a2) for the LO port in question. The
trace should be -15
±
2 dBm across the band as displayed on the
HP 8510 CRT. If this is the case the LO ALC is adjusted properly.
If the message
NO IF FOUND
is displayed and/or a trace is not observed
at the level mentioned, the LO ALC adjustment needs to be made,
continue with step 3.
NOTE
Step 3 should only be performed when an amplifier, coupler, or detector has
been replaced.
3. Connect the oscilloscope and negative detector as shown in Figure 6-16
on page 6-32. Turn the gain adjustment fully clockwise. Oscillations
should be visible on the oscilloscope. Turn the gain adjustment counter
clockwise until the oscillations just disappear. Recheck the user
parameters in step 2 and then continue with step 4.
4. Choose a power meter that covers the 4 to 6 GHz frequency range
(HP 436A meter and HP 8485A sensor). Calibrate the power meter
and attach a 3.5 mm 10 dB fixed attenuator (HP 8493C) and an adapter
(female to female) to the sensor. Connect the sensor/attenuator to the LO
output of the port in question. Turn the LO level adjustment so that
+12 dBm is displayed on the power meter. The LO output is now set at
+22 dBm. Select {
SINGLE POINT MODE
} from the [
STIMULUS
] menu, then
using the RPG knob, slowly scan over the frequency band.
5. Adjust the LEVEL pot over the full adjustment range to make sure there
are no oscillations at any power level. Then turn the GAIN pot three
additional turns.
Содержание 85106D
Страница 11: ...x HP 85106D mm Wave System ...
Страница 95: ...4 2 HP 85106D mm Wave System Service and Troubleshooting Figure 4 1 HP 85106D System Block Diagram ...
Страница 96: ...HP 85106D mm Wave System 4 3 Service and Troubleshooting Figure 4 2 HP 85106D Option 001 System Block Diagram ...
Страница 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Страница 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 149: ...5 36 HP 85106D mm Wave System HP 85104A Test Set Module Assembly Procedure Figure 5 12 HP 85104A Assembly Procedure ...
Страница 153: ...5 40 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 13 HP Q85104A Replaceable Parts ...
Страница 156: ...HP 85106D mm Wave System 5 43 HP 85104A Test Set Module Replaceable Parts Figure 5 14 HP U85104A Replaceable Parts ...
Страница 159: ...5 46 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 15 HP V85104A Replaceable Parts ...
Страница 162: ...HP 85106D mm Wave System 5 49 HP 85104A Test Set Module Replaceable Parts Figure 5 16 HP W85104A Replaceable Parts ...
Страница 208: ......
Страница 212: ...6 48 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 28 HP 85105A Major Assemblies ...
Страница 214: ...6 50 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 29 HP 85105A Semi rigid RF Cables ...
Страница 218: ...6 54 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 31 Wire Harnesses ...
Страница 220: ...6 56 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 32 HP 85105A Front Panel ...
Страница 222: ...6 58 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 33 Rear Panel Standard Option 050 ...
Страница 228: ...6 64 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 37 HP 85105A Accessories Standard ...
Страница 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Страница 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Страница 244: ... vi HP 85106D mm Wave System ...