HP 85106D mm-Wave System
1-11
Getting Started
Preparing the Site
System Voltages
Available
All HP 85106D system instruments must be set to the same voltage as the
system rack; either 120 Vac or 220 Vac. The system instruments are set at
the factory to 120 Vac, with the exception of the Option 230 system, which
is set to 220 Vac.
CAUTION
System cabinet fans may be permanently damaged if a 120 V system is
plugged into a 230 Vac power outlet. Cabinet fans are wired for either 120 V
or 230 V, not both. Therefore, a system wired for 120 V operation cannot be
switched to 230 V operation by changing only individual instrument voltage
selection switches.
Consult individual instrument manuals to change voltages from 120 V to
100 V (120 V systems only), or from 220 V to 240 V (Option 230 system, or
230 V systems only).
Table 1-6
Maximum VA Ratings and BTU/hour Ratings of HP Instruments
Instrument
Maximum
VA Rating
1
Actual VA
Subtotal
Maximum
BTU/hour
Actual
BTU/hour
Subtotal
Standard Equipment
HP 85101 Display Processor
250
___________
850
___________
HP 85102 IF Detector
210
___________
714
___________
HP 85105 Test Set Controller
270
___________
918
___________
HP 8360 Synthesized Source
(quantity of two)
400
___________
1360
___________
HP 8517 Coaxial Test Set
(HP 85106D Option 001
with or without Option 007)
145
___________
323
___________
Totals:
Standard System
1505 VA
___________
5117
___________
Option 001.002
1675 VA
___________
5525
___________
1. Values are based on 120 Vac supplied to each instrument at 60 Hz.
Содержание 85106D
Страница 11: ...x HP 85106D mm Wave System ...
Страница 95: ...4 2 HP 85106D mm Wave System Service and Troubleshooting Figure 4 1 HP 85106D System Block Diagram ...
Страница 96: ...HP 85106D mm Wave System 4 3 Service and Troubleshooting Figure 4 2 HP 85106D Option 001 System Block Diagram ...
Страница 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Страница 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 149: ...5 36 HP 85106D mm Wave System HP 85104A Test Set Module Assembly Procedure Figure 5 12 HP 85104A Assembly Procedure ...
Страница 153: ...5 40 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 13 HP Q85104A Replaceable Parts ...
Страница 156: ...HP 85106D mm Wave System 5 43 HP 85104A Test Set Module Replaceable Parts Figure 5 14 HP U85104A Replaceable Parts ...
Страница 159: ...5 46 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 15 HP V85104A Replaceable Parts ...
Страница 162: ...HP 85106D mm Wave System 5 49 HP 85104A Test Set Module Replaceable Parts Figure 5 16 HP W85104A Replaceable Parts ...
Страница 208: ......
Страница 212: ...6 48 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 28 HP 85105A Major Assemblies ...
Страница 214: ...6 50 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 29 HP 85105A Semi rigid RF Cables ...
Страница 218: ...6 54 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 31 Wire Harnesses ...
Страница 220: ...6 56 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 32 HP 85105A Front Panel ...
Страница 222: ...6 58 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 33 Rear Panel Standard Option 050 ...
Страница 228: ...6 64 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 37 HP 85105A Accessories Standard ...
Страница 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Страница 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Страница 244: ... vi HP 85106D mm Wave System ...