2-8
HP 85106D mm-Wave System
Operation
System Set Up
CAUTION
Do not 27 dBm incident at Port 2 or Port 1. System specifications
are determined with about 0 dBm incident at Port 2 or Port 1. Greater than
+ 27 dBm damages the test set making it necessary to replace expensive
parts. Observe proper precautions, especially when measuring amplifiers
having greater than about 20 dB of gain.
Figure 2-2
Typical Port 1 and Port 2 Configuration
1. To align the test ports for connection, raise the modules to the same
height by fully extending the four feet on each stand. If the surface is
level the test set modules should not rock. If either module is not stable
make the final adjustment by changing the height of the feet using the
height-adjustment screws.
2. Release the stand lock by moving the lever, then center the module
within the range of motion. Lock the slider. Small movements of the test
port (up to 1/2-inch) can be made by turning the small knob at the front
of the stand. Typically, Port 1 is fixed in position and Port 2 is moved
toward it to make the connection.
WARNING
Never look directly into the open end of a waveguide section while RF source
power is on. Although the signal levels in this setup are relatively low,
mm-wave energy is non-ionizing radiation that can be extremely harmful to the
eye. The results may be eye damage or permanent blindness.
Содержание 85106D
Страница 11: ...x HP 85106D mm Wave System ...
Страница 95: ...4 2 HP 85106D mm Wave System Service and Troubleshooting Figure 4 1 HP 85106D System Block Diagram ...
Страница 96: ...HP 85106D mm Wave System 4 3 Service and Troubleshooting Figure 4 2 HP 85106D Option 001 System Block Diagram ...
Страница 117: ...4 24 HP 85106D mm Wave System Service and Troubleshooting Check Unratioed Power Levels ...
Страница 139: ...5 22 HP 85106D mm Wave System HP 85104A Test Set Module Troubleshooting Procedures ...
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 149: ...5 36 HP 85106D mm Wave System HP 85104A Test Set Module Assembly Procedure Figure 5 12 HP 85104A Assembly Procedure ...
Страница 153: ...5 40 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 13 HP Q85104A Replaceable Parts ...
Страница 156: ...HP 85106D mm Wave System 5 43 HP 85104A Test Set Module Replaceable Parts Figure 5 14 HP U85104A Replaceable Parts ...
Страница 159: ...5 46 HP 85106D mm Wave System HP 85104A Test Set Module Replaceable Parts Figure 5 15 HP V85104A Replaceable Parts ...
Страница 162: ...HP 85106D mm Wave System 5 49 HP 85104A Test Set Module Replaceable Parts Figure 5 16 HP W85104A Replaceable Parts ...
Страница 208: ......
Страница 212: ...6 48 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 28 HP 85105A Major Assemblies ...
Страница 214: ...6 50 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 29 HP 85105A Semi rigid RF Cables ...
Страница 218: ...6 54 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 31 Wire Harnesses ...
Страница 220: ...6 56 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 32 HP 85105A Front Panel ...
Страница 222: ...6 58 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 33 Rear Panel Standard Option 050 ...
Страница 228: ...6 64 HP 85106D mm Wave System HP 85105A mm Wave Controller Replaceable Parts Figure 6 37 HP 85105A Accessories Standard ...
Страница 238: ...B 2 HP 85106D mm Wave System Upgrade Paths ...
Страница 243: ...HP 85106D mm Wave System v work surface installation 1 12 ...
Страница 244: ... vi HP 85106D mm Wave System ...