5
may
be
dierent
from
the
value
set
for
the
output.
If
these
two
gures
are
not
the
same
,
it
is
indicated
in
the
ST
A
Tus:OPERation
register
.
Attenuator
If
you
have
the
optional
attenuator
installed,
the
value
returned
for
the
power
is
dependent
on
whether
you
are
in
power
or
attenuation
mode
(see
\[:SOURce]:POW
er:A
TT
enuation:A
UTO?").
If
you
are
in
power
mode
,
the
value
returned
is
the
output
power
.
If
you
are
in
attenuation
mode
,
the
value
returned
is
the
laser
output
power
,
and
you
must
also
use
the
attenuation
value
to
calculate
the
output
power
(see
\[:SOURce]:POW
er:A
TT
enuation?").
If
you
specify
MIN ,
DEF ,
or
MAX
with
the
query
,
it
will
return,
respectively
,
the
minimum,
default,
or
maximum
amplitude
level.
Example
OUTPUT
724;":POW:UNIT
DBM"
OUTPUT
724;":POW?
MAX"
ENTER
724;A$
OUTPUT
724;":POW
",A$
[:SOURce]:PO
W
er:UNIT
Syntax
[:SOURce ]:POWer:UNIT
<wsp>
DBM j DBMW jWatt
Description
This
command
sets
the
power
units
.
Set
the
units
to
dBm
by
sending
DBM
or
DBMW .
Set
the
units
to
watts
by
sending
W .
The
default
units
are
dBm.
[:SOURce]:PO
W
er:UNIT?
Syntax
[:SOURce ]:POWer:UNIT?
Description
The
query
returns
the
units
currently
being
used.
The
response
is
either
0
for
dBm
or
dBmW
,
or
2
for
W
atts
.
Example
OUTPUT
724;":POW:UNIT
DBM"
OUTPUT
724;":POW:UNIT?"
ENTER
724;A$
5-26
Remote
Commands
Содержание 8167A
Страница 2: ......
Страница 4: ......
Страница 26: ......
Страница 110: ......
Страница 130: ......
Страница 150: ......
Страница 164: ...D Figure D 1 Test Setup for the Wavelength Tests D 4 Performance Tests ...
Страница 168: ...D Figure D 2 Test Setup for the Maximum Output Power Test except HP 8168F D 8 Performance Tests ...
Страница 177: ...D Figure D 5 Test Setup for the Source Spontaneous Emission Test Performance Tests D 17 ...
Страница 240: ......