6
Example
5
-
Increased
Tuning
Linearity
18
pm,
which
is
larger
than
the
1
pm
tolerance
allowed
in
this
example
.
The
measured
wavelength
reading
is
therefore
sent
to
the
TLS
using
the
HP-IB
command
WAVEACT .
The
TLS
will
now
detune
itself
to
reduce
the
dierence
.
The
new
wavelength
is
read
with
the
wavelength
meter
,
and
the
correction
loop
repeated
until
the
dierence
is
reduced
to
the
allowed
tolerance
.
Once
the
correct
wavelength
is
achieved,
the
receiver
can
be
read.
This
process
is
repeated
for
each
wavelength
in
the
sweep
(Figure
6-3 ).
Figure
6-3.
Flow
chart
describing
Increased
Tuning
Linearity
program.
In
this
example
,
the
allowed
wavelength
dierence
is
1
pm,
that
is
the
minimum
achievable
on
the
TLS.
This
will
typically
take
several
cycles
through
the
correction
loop
.
Allowing
a
larger
tolerance
will
reduce
the
average
number
of
correction
cycles
and
overall
test
time
.
Programming
Examples
6-13
Содержание 8167A
Страница 2: ......
Страница 4: ......
Страница 26: ......
Страница 110: ......
Страница 130: ......
Страница 150: ......
Страница 164: ...D Figure D 1 Test Setup for the Wavelength Tests D 4 Performance Tests ...
Страница 168: ...D Figure D 2 Test Setup for the Maximum Output Power Test except HP 8168F D 8 Performance Tests ...
Страница 177: ...D Figure D 5 Test Setup for the Source Spontaneous Emission Test Performance Tests D 17 ...
Страница 240: ......