5
Figure
5-2.
The
Status
Registers
The
OPERation
branch
of
the
status
registers
indicates
things
that
can
happen
during
normal
operation.
The
QUEStionable
branch
indicates
error
conditions
.
:ST
A
Tus:OPERation:CONDition?
Syntax
:STATus:OPERation:CONDition?
Description
This
query
reads
the
contents
of
the
OPERation:CONDition
register
.
Only
two
bits
of
the
condition
register
are
used:
Bit
8,
which
is
1
when
the
output
power
that
you
set
exceeds
what
the
laser
is
capable
of
producing.
Bit
9,
which
is
1
while
the
instrument
is
performing
its
power-up
initialization.
Example
OUTPUT
724;":STAT:OPER:COND?"
ENTER
724;A$
5-30
Remote
Commands