1
e.
Press
NNNNN
#
to
skip
over
the
dwell
parameter
.
f.
Type
in
1
on
the
keypad
and
press
4
ENTER
5
to
set
the
number
of
times
that
the
sweep
is
to
be
performed.
13.
On
the
laser
source
,
press
NNNNNNNNNNNNNNNNNNNNNN
Pmax ! P
to
set
the
output
power
to
the
highest
value
that
can
be
maintained
for
the
full
sweep
range
.
14.
On
the
power
meter:
a.
Set
to
1535nm
(4
Pa
ram
5 ! !1535.0nm).
b.
Make
sure
that
the
parameter
cursor
is
in
the
units
position
(that
is
,
at
the
second
\5").
P
erforming
the
W
avelength
Characteristic
Measurement
15.
Make
sure
that
the
laser
source
is
still
active
.
16.
On
the
laser
source
,
press
NNNNNNNNNNNNNNNNNNNN
Manual
.
17.
Read
the
value
for
the
power
at
1535nm
on
the
multimeter
.
18.
Repeat
the
following
steps
at
each
wavelength
in
the
sweep
range
a.
On
the
laser
source
,
press
NNNNNNNNNNNNNN
Next
.
b.
On
the
multimeter
,
increment
(press
4
*
5 ).
c.
Read
the
value
for
the
power
at
the
wavelength
on
the
multimeter
.
19.
On
the
laser
source
,
press
NNNNNNNNNNNNNN
Stop
to
end
the
sweep
.
This
procedure
can
be
used
in
practice
to
measure
the
wavelength
characteristic
of
a
component.
1-12
Getting
Started
Содержание 8167A
Страница 2: ......
Страница 4: ......
Страница 26: ......
Страница 110: ......
Страница 130: ......
Страница 150: ......
Страница 164: ...D Figure D 1 Test Setup for the Wavelength Tests D 4 Performance Tests ...
Страница 168: ...D Figure D 2 Test Setup for the Maximum Output Power Test except HP 8168F D 8 Performance Tests ...
Страница 177: ...D Figure D 5 Test Setup for the Source Spontaneous Emission Test Performance Tests D 17 ...
Страница 240: ......