
7
Calibrate the Logic Trigger of the Self Cal.
a
Obtain a BNC 50-
Ω
, 9-inch cable.
b
Touch the Procedure field, then touch Logic Trigger from the pop-up
menu.
c
Touch the Start field and follow the instructions on the display.
d
After completion of the Logic Trigger calibration, remove the cable
from the instrument.
8
Calibrate Ext Trig Null of the Self Cal.
a
Connect a BNC 50-
Ω
, 9-inch cable to the BNC tee adapter. Connect
the BNC 50
Ω
(f)/SMB(m) adapter to one end of the BNC tee adapter
using a BNC(m)(m) adapter. Once you touch Start, the instrument
will prompt you to connect the cables to the appropriate locations on
the rear panel of the instrument.
b
Touch the Procedure field, then touch Ext Trig Null from the pop-up
menu.
c
Touch the Start field and follow the instructions on the display.
d
After completion of External Trigger Null Calibration, remove the
cables from the instrument.
Protect the operational accuracy calibration factors
1
Remove power from the instrument. Pull out halfway all cards and
filler panels located above the oscilloscope, then remove the filler
panels and cards located above the oscilloscope.
2
Remove the HP 16533/34A module and set the
PROTECT/UNPROTECT switch to the PROTECT position. On a
multi-card module, the PROTECT/UNPROTECT switch on all cards
must be set to PROTECT.
3
Reinstall the modules and filler panels.
Installation and Operator’s Service
Protect the operational accuracy calibration factors
11–15
Содержание 16533A
Страница 6: ...vi ...
Страница 11: ...1 Run Stop Options ...
Страница 15: ...2 Autoscale ...
Страница 21: ...3 The Channel Menu ...
Страница 23: ...Channel Field Menu Map The Channel Menu 3 3 ...
Страница 29: ...4 Time Base Functions ...
Страница 42: ...4 14 ...
Страница 43: ...5 The Marker Menu ...
Страница 45: ...Markers Field Menu Map The Marker Menu 5 3 ...
Страница 46: ...Markers Field Menu Map Continued The Marker Menu 5 4 ...
Страница 65: ...6 The Trigger Menu ...
Страница 67: ...Trigger Field Menu Map The Trigger Menu 6 3 ...
Страница 84: ...6 20 ...
Страница 85: ...7 The Display Menu ...
Страница 87: ...Display Field Menu Map The Display Menu 7 3 ...
Страница 94: ...7 10 ...
Страница 95: ...8 The Auto Measure Menu ...
Страница 97: ...Auto Measure Menu Map The Auto Measure Menu 8 3 ...
Страница 105: ...9 The Waveform Selection Menu ...
Страница 107: ...Waveform Selection Menu Map The Waveform Selection Menu 9 3 ...
Страница 116: ...9 12 ...
Страница 117: ...10 Specifications and Characteristics ...
Страница 125: ...11 Installation and Operator s Service ...
Страница 140: ...11 16 ...
Страница 148: ...Glossary 4 ...