38
Chapter 6
Configuration / Setting
Subject to technical changes
Rev. 1.01.08
[P092]t2.d1 - Mode of Operation of Timer 2, Defini-
tion 1
Ø
[P090]t1.d1 - Mode of Operation of Timer 1, Defini-
tion 1
[P093]t2.d2 - Mode of Operation of Timer 2, Defini-
tion 2
Ø
[P091]t1.d2 - Mode of Operation of Timer 1, Defini-
tion 2
[P094]t3.d1 - Mode of Operation of Timer 3, Defini-
tion 1
Ø
[P090]t1.d1 - Mode of Operation of Timer 1, Defini-
tion 1
[P095]t3.d2 - Mode of Operation of Timer 3, Defini-
tion 2
Ø
[P091]t1.d2 - Mode of Operation of Timer 1, Defini-
tion 2
[P096]t4.d1 - Mode of Operation of Timer 4, Defini-
tion 1
Ø
[P090]t1.d1 - Mode of Operation of Timer 1, Defini-
tion 1
[P097]t4.d2 - Mode of Operation of Timer 4, Defini-
tion 2
Ø
[P091]t1.d2 - Mode of Operation of Timer 1, Defini-
tion 2
[P098]t.rES - Mode of Operation of Timer after Soft-
Reset
A soft reset stands for a sensor break
,
a reset of a set
point to zero or a passivation of a zone.
6.10 Configuration of Operation
[P099]Edit - Entry Mode Directly
[P100]iLoc - Entry Lock on Info Level
[P110]ULoc - Entry Lock on User/System level
After restart of the device
,
all data entry except setpoint
value and degree of operation is automatically blocked.
AL.t1
Timer is started when the actual value reaches
the 5K band around the set point directly after
start of the controller.
After termination: Start of timer 1.
AL.t2
Timer is started when the actual value reaches
the 5K band around the set point directly after
start of the controller.
After termination: Start of timer 2.
AL.t3
Timer is started when the actual value reaches
the 5K band around the set point directly after
start of the controller.
After termination: Start of timer 3.
AL.t4
Timer is started when the actual value reaches
the 5K band around the set point directly after
start of the controller.
After termination: Start of timer 4.
L.t1
Timer is started when the actual value reaches
the 5K band around the set point.
After termination: Start of timer 1.
L.t2
Timer is started when the actual value reaches
the 5K band around the set point.
After termination: Start of timer 2.
L.t3
Timer is started when the actual value reaches
the 5K band around the set point.
After termination: Start of timer 3.
L.t4
Timer is started when the actual value reaches
the 5K band around the set point.
After termination: Start of timer 4.
AL
Timer is started when the actual value reaches
the 5K band around the set point directly after
start of the controller.
L
Timer is started when the actual value reaches
the 5K band around the set point.
AS
Timer is started when set point > 100 °C and
actual value < 90 °C.
AS.t1
Timer is started when set point > 100 °C and
actual value < 90 °C.
After termination: Start of timer 1.
AS.t2
Timer is started when set point > 100 °C and
actual value < 90 °C.
After termination: Start of timer 2.
AS.t3
Timer is started when set point > 100 °C and
actual value < 90 °C.
After termination: Start of timer 3.
AS.t4
Timer is started when set point > 100 °C and
actual value < 90 °C.
After termination: Start of timer 4.
[run]
Timer keeps on running.
StoP
Timer is stopped and reset.
Auto
Timer is stopped
,
reset and after soft reset
started again in case of t*.d2 = Auto and/or
A.**.
[on]
After selection the parameters in operator and
system level can
directly
be changed by key--
pad.
In this mode the direct data entry for degree of
operation/ set point value by decrease/in-
crea-se button is feasible.
oFF
After selection the parameters in operator and
system level can only be changed after un--
blocking by keypad by decrease/increase but--
ton.
on
No data entry on information level feasible.
[oFF]
Data entry on information level feasible.