Residual chlorine monitor HR-200 Operation procedure 6
(R
ema
rks
, c
o
d
e
erro
rs
)
Measurement
s exceeding
the measurable range
Th
e mea
sur
ed
value
for
resid
ual chlor
in
e b
lin
ks whe
n it falls
out
side
the
measura
ble r
ang
e (3
.0
0 mg/
L
max.;
5.00
mg/L
ma
x. w
hen
r
ange
cut is n
ot selecte
d).
Ta
ke remed
ia
l actio
n as instr
ucted
be
low
.
Symp
to
m
P
ossib
le cause
R
emed
ial action
Blin
king d
ispla
y
(out of
the
measura
ble
rang
e)
T
he con
ce
ntration is hig
h.
Che
ck
resid
ual
chl
orine
con
centration
wi
th
DPD
teste
r first,
cont
act
us if
sp
a
n calib
rati
on is unsucce
ssful
un
der th
e normal ran
ge.
T
he tempera
ture
sen
sor
is
ab
normal
.
Pe
rfo
rm th
e te
mp
er
at
ur
e cal
ib
ra
tion
.
T
he sen
so
r
is
no
t
co
rre
ctly conn
ected.
Che
ck the wiri
ng and
p
ro
per
ly conn
ect
the le
ad wires.
Cont
ac
t us if the syst
em
is not recovered.
Error codes
Description o
f error codes
Displa
yDispla
y
Priority
level of
error
*1
Err
o
r
D
escription
Oc
casion
E-12
SENSOR
ERROR
8
S
pan
cal
ibratio
n error
T
he senso
r
curren
t is low du
ring the sp
an
cal
ibratio
n.
Durin
g ca
lib
ra
tion
*2
E-13
Z
ero calib
ration error
T
he senso
r
curren
t is hig
h durin
g
the
ze
ro
ca
libratio
n.
E-21
4
Tempe
ratu
re se
nsor discon
nection
e
rro
r
T
he
le
ad
wire
from the
tempera
ture
sen
sor i
s o
pene
d.
Durin
g me
asuremen
t or calib
ration
E-22
4
Tempe
ratu
re se
nsor short-circuit
e
rro
r
T
he le
ad wir
e
from the
tempera
ture
sen
sor i
s sho
rt-circui
te
d.
E-25
7
Tempe
ratu
re me
asurab
le rang
e
e
rro
r
T
he measu
re
d te
mp
erature valu
e
is
u
nder
-5
C or over 55
C.
E-27
9
Tempe
ratu
re ca
lib
ra
tion error
T
he m
easu
re
d
te
mp
erature
valu
e i
s u
nder -5
C or over 55
C.
Durin
g ca
lib
ra
tion
E-61
5
F
low rate error
L
ow fl
ow ra
te
is
che
cke
d by
a
flo
w
swi
tch
.
Durin
g me
asuremen
t or calib
ration
*3
E-75
6
S
en
sor
e
rro
r
T
he curren
t sign
al th
rough
A and
K is to
o high
.
D
urin
g me
asuremen
t or calib
ration
E-90
SYSTEM
ERROR
1
System error
In
te
rn
al
com
m
un
ic
at
io
n err
o
r
At any
time sys
tem error
E-91
2
A
/D converter ma
lfunction
E-92
3
T
he stored dat
a, such as
se
tting
s
a
nd cali
bration val
ues,
is
vol
ati
lized
.
*1:
A
hi
gher prio
rity error code
wi
ll be di
splaye
d
w
hen multip
le errors occur
.
*2:
It
is
p
ossible
to cali
brate sensor with E-12 or E-1
3.
Ho
wever
, there
is
no
wa
y to
cali
brate
th
e
se
nsor with E-2
1 to
E-27
. T
ake remedi
al action in
a
cco
rdance
wi
th
“ Remed
ial
actions for
error codes
” (p
age
64).
*3:
F
or
E-61, the
w
arnin
g i
s issu
ed
onl
y wh
en
th
e
flow
switch i
s
in
st
a
lled
a
nd th
e no-flow al
arm
is checke
d
o
n th
e co
nverter
.
h19136 | UTC 2020/03/30 3:34:51
CONFIDENTIAL