Troubleshooting
62
Measurements exceeding the measurable range
The measured value for residual chlorine blinks when it falls outside the measurable range
(3.00 mg/L max.; 5.00 mg/L max. when range cut is not selected).
Take remedial action as instructed below.
An abnormal
value is
displayed.
Unexpected short circuit or
open circuit in the sensor
cable.
Check the sensor cable.
Loosened terminal
connection
Check the terminal block.
Unexpected
operation
The circuit board is
corroded by corrosive gas.
Supply clean instrumentation air from the air purge inlet.
Power
supply error
The power supply voltage
falls outside the rated
voltage range.
Check the power supply voltage.
Something wrong with
wiring blown fuse
Check the power wiring.
Internal trouble
Contact us.
Symptom
Possible cause
Remedial action
Symptom
Possible cause
Remedial action
Blinking display
(out of the
measurable
range)
The concentration is high.
Check residual chlorine concentration with DPD tester first,
contact us if span calibration is unsuccessful under the
normal range.
The temperature sensor is abnormal.
Perform the temperature calibration.
The sensor is not correctly connected.
Check the wiring and properly connect the lead wires.
Contact us if the system is not recovered.