Data
69
Sensor
Product name
Polarographic residual chlorine sensor
Model
Model
RA-10
RA-20
Sampling
type
Overflow type
Inline type
Measuring method
Polarography
Measuring object
Free chlorine concentration in the sample water
Measurement range
0.00 mg/L to 3.00 mg/L
Temperature compensation
Automatic temperature compensation using platinum RTD (Pt1000).
Performance
Linearity
Within
5
of the full scale
Repeatability Within
3
of the full scale
Response
time
Within 60 seconds to 90
(sample change at the inlet)
Zero drift
Within
1
of the full scale per month
Span drift
Within
10
of the full scale per month
Sample
condition
1
Temperature 0
C to 45
C (avoid freezing)
Flow rate
1.3 L/min to 2.0 L/min
0.6 L/min to 1.0 L/min constant
2
Pressure
Within 0.5 MPa
3
pH
5.8 pH to 8.6 pH constant
4
Electrical
conductivity
More than 10mS/m
5
Calibration
Zero
calibration
Chlorine calibration zero liquid
Span
calibration
Span calibration (comparing to the measurement value of DPD method, including
electric zero calibration)
Wetted materials
PVC, PPO, FKM, EPDM
Electrode material
Au, AgCl, C
Bead material
SiO
2
or ceramic
Filter material
Nylon
Pipe
connection
Sample inlet PREFAB JOINT TS16A (ASHAHI)
Sample
outlet
PREFAB JOINT TS16A (ASHAHI)
6
Cleaning method
Mechanical polishing with bead, electrochemical cleaning
7
Cable length
Standard: 2 m (max. extension: 40 m)
Operating temperature range 0
C to 45
C
8
Storage temperature
0
C to 45
C
External dimensions
221 (W)
278 (H)
201 (D) mm
9
221 (W)
157 (H)
201 (D) mm
9
Mass
Approx. 2.3 kg
Approx. 1.8 kg
Standard accessory
Cleaning bead
10
, mesh filter, sanding paper, O-ring (S50, S30, P14, P11)
1:
Oxidants such as ozone, hydrogen peroxide, bromine and reductants such as sodium thiosulphates, sodium
sulfite will affect measured data.
2:
Sample flow must be kept constant. The sensor is affected by flow rate.
3:
If the regulator recommended by HORIBA Advanced Techno is used inlet pressure should be over 0.1 MPa.
4:
Sample pH should be constant. This sensor is dependent on the pH which effects the dissociation of
hypochlorous acid.
5:
The conductivity must be more than 10 mS/m (100
S/cm).