Measurement
31
When “CtrL”,
“AL”, or “t” is
selected.
Relay setting condition
screen (the numeral at
the lower left indicates
the relay number)
-
Control value setting
screen
Displays the set value (residual
chlorine concentration or
temperature). The set value can be
changed.
RC
: 0.00 to
3.00 (mg/L)
Temperature:
0.0 to 100.0
(
C)
When “CtrL” is
selected.
Control width value
screen
Displays or is used to specify the
control width using the residual
chlorine concentration value.
0.02 to 1.00
(mg/L)
When “AL” or t is
selected.
Delay time screen
Displays or is used to specify the
delay time in seconds.
0 to 600 (s)
When “HoLd” is
selected.
Relay setting condition
screen (the numeral at
the lower left indicates
the relay number)
-
Delay time screen
Displays or is used to specify the
delay time in seconds.
0 to 600 (s)
When “CLn” is
selected.
Output during cleaning
screen
Shows that data is output during
cleaning and during HOLD after
cleaning.
-
Cleaning check screen
Cleaning check screen
-
Wait time before starting
the automatic cleaning
Displays the time before starting the
next cleaning in a format of hours-
minutes-seconds.
Note:
When the wait time is 10 h or more, it
is displayed in the format of
hours.minutes.
-
-
Cleaning test screen
Performs cleaning while ENT is
held down.
Holds the output during the HOLD
time after cleaning when ENT is
released.
-
Calibration data check screen
Calibration data check screen
No calibration
data
Screen for no
calibration data
Indicates that there is no previous
calibration data.
-
Display
Display condition
Item
Description
Setting range