Installation
12
Installation
Installation environment
In order to ensure the stable use of the HR-200, install it in a location that meets the following
requirements.
Converter
The location should be well ventilated.
The ambient temperature should be between
20
C and 55
C.
The HR-200 should not be exposed to direct sunlight.
Hot air at the location should be minimal.
The HR-200 should not be exposed to direct high radiation heat.
The relative humidity should be between 5
and 90
.
The HR-200 should not be exposed to splashed water or chemical agent.
The mechanical vibration in the location should be minimal.
The location should have adequate space to allow maintenance and wire connecting
work.
The location should be free from dust and corrosive gas.
The location should be minimally affected by an electromagnetic fields.
The altitude should be no more than 2000 m.
The power supply voltage should fall within the range from 100 V to 240 V AC
10
.
Sensor
The location should have adequate space to allow for check and maintenance of the
sensor.
The ambient temperature should be between 0
C and 45
C.
Spilled water should not be a serious matter.
The sample water should be free from air bubbles.
The sample should not corrode the wetted part of the sensor.
The sample water should not be frozen.
The HR-200 should not be exposed to direct sunlight.