![horiba APMA-370 Скачать руководство пользователя страница 75](http://html1.mh-extra.com/html/horiba/apma-370/apma-370_operation-manual_2162193075.webp)
6 FUNCTIONALITIES
63
6.6.4
Data saving
To save data in the memory manually, press the [SAVE] button on the MENU/SYSTEM
screen. Before turning OFF the APMA-370, be sure to perform this operation to save the data
that has not yet been saved in the memory.
Note
The average and integration values are stored in the flash memory every 10 minutes.
Before turning OFF the power, be sure to save those data in the memory.
If accidental power outage occurs, data during a period of 10 minutes maximum may not be
recorded.
1.
Press the [SAVE] button on the MENU/SYSTEM screen. The message confirming
the saving of the data will appear.
Fig. 80 Message confirming the saving of the data
The buttons allow you to perform the following operations.
2.
Press the [YES] button. The message of saving will be displayed and the data will
start being saved. After the saving of the data is completed, the MENU/SYSTEM
screen is displayed again.
Fig. 81 Message of saving
[YES]:
Saves the data in the memory.
[NO]:
Saves no data to the memory and returns to the MENU/SYSTEM screen.