4 CALIBRATION
24
CAL.
The following CAL. screen for setting will appear.
Fig. 30 CAL. screen (for SPAN)
5.
Change the setting by entering time on the time setting screen or pressing either
[YES] or [NO] button on the CAL. screen, and then press the [SET] key.
The setting will be changed and the AIC SEQUENCE screen will be displayed again.
Note
Any process for which time is set to 0 min is skipped and the AIC sequence proceeds to the next
step. For example, if WAIT for SPAN is set to 0 min, no span gas will be supplied.
If CAL. is set to NO, calibration is not performed in the AIC sequence.
If the total time of the AIC sequence exceeds the value of AIC INTERVAL minus 10 min, the AIC
INTERVAL setting will be automatically changed to the value of the total time of the AIC sequence
plus 10 min.
6.
Press the [CLOSE] button on the AIC SEQUENCE screen.
The MENU screen is displayed again.
An example of the AIC sequence is shown in the following diagram:
Fig. 31 An example of the AIC sequence