4 CALIBRATION
18
AIC MODE
Specify the method of starting the AIC.
Pressing the displayed AIC MODE setting will display the AIC MODE screen.
Fig. 21 AIC MODE screen
Press the button of the item to be set.
Note
Manual AIC start and the start via the RS-232C port are valid regardless of this setting.
If an AIC start signal is inputted externally while an AIC sequence is in progress, this signal will
be disregarded and the ongoing AIC sequence will be continued.
Tip
For the telemeter connection specifications, to execute AIC using the internally set START TIME and
INTERVAL automatically even if the start signal is not inputted because of telemeter malfunction, set
AIC MODE to EXTERNAL.
Item
Description
INTERNAL
Selects the mode of using the internal clock to execute AIC at the specified start time and intervals with.
EXTERNAL
Selects the mode of using the external start signal (external contact input) to start AIC.
For the telemeter connection specifications, if the telemeter input contact is open (telemeter malfunction),
AIC will be started using the internal clock.
OFF
Selects the mode without AIC automatic start.