6 FUNCTIONALITIES
44
6.3
Maintenance Menu
Fig. 55 MENU/MAINTENANCE screen
The keys allow you to perform the following operations.
6.3.1
Analog output
Press the [ANALOG OUTPUT] button on the MENU/MAINTENANCE screen. The ANALOG
OUTPUT screen will be displayed.
This screen allows you to check and control the analog output.
Fig. 56 ANALOG OUTPUT screen
The current output modes of ANALOG OUTPUT 1 (momentary value) and ANALOG OUTPUT
2 (which varies depending on the specification; integration value for the standard specifica-
tion) are displayed as buttons.
Note
All outputs are always in the [MEAS.] mode (the current measured value is being outputted) first.
[ANALOG OUTPUT]:
Displays the ANALOG OUTPUT screen (Fig. 56 on page 44).
[ANALOG INPUT]:
Displays the ANALOG INPUT screen (Fig. 59 on page 48).
[HOUR METER]:
Displays the HOUR METER screen (Fig. 60 on page 49).
[MEAS.]:
The current measured value is being outputted. This is normal mode.
[XX%]:
XX% of the full scale is being outputted. The settable value is between 0%
(example: about 0 V) and 100% (example: about 1 V) in steps of 10%.