4 CALIBRATION
29
4.4.4
Finishing calibration
1.
When using the calibration gas line, display the MODE screen and change the
measurement line to [MEAS.].
When using the measured gas line, supply the sample gas to the measured gas
line.
2.
Press the [CLOSE] key on the CAL. screen. The MEAS. screen will be displayed
again and the measurement will start.