
6 FUNCTIONALITIES
45
The buttons allow you to select and check the output modes (see below).
The keys allow you to perform the following operations.
Checking output
1.
On the ANALOG OUTPUT screen, press the button of the output mode to be
changed.
With every pressing of the button, the button display and the actual output mode
are automatically changed in the following order:
2.
To finish output check, press the [CLOSE] key to return to the MENU/
MAINTENANCE screen.
Note
Pressing the [CLOSE] key will put all the output back to the measured values.
[CLOSE]:
Returns to the MENU/MAINTENANCE screen.
[ZERO]:
Displays the DA ADJUST/ZERO screen for zero calibration of analog output
(Fig. 57 on page 46).
[SPAN]:
Displays the DA ADJUST/SPAN screen for span calibration of analog output
(Fig. 58 on page 47).
[OUTPUT]:
Outputs the selected output via the analog output terminal.
[MEAS.]
Measured value output
->
[0%]
0% output
->
[10%]
10% output
->
[20%]
20% output
↑
↓
[100%]
100% output
[30%]
30% output
↑
↓
[90%]
90% output
[40%]
40% output
↑
↓
[80%]
80% output
<-
[70%]
70% output
<-
[60%]
60% output
<-
[50%]
50% output
Содержание APMA-370
Страница 1: ...Ambient CO monitor APMA 370 Operation Manual eng CODE GZ0000051249J...
Страница 8: ......
Страница 12: ......
Страница 106: ......
Страница 107: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...