4 CALIBRATION
23
4.3.3
Setting the AIC sequence
To set the AIC sequence, go to the AIC SEQUENCE screen.
1.
Press the [MENU] key on the MEAS. screen.
2.
Press the [
] or [
] key to display the MENU/SETTING screen.
Fig. 27 MENU/SETTING screen
3.
Press the [AIC SEQUENCE] button. The AIC SEQUENCE screen will be displayed.
Fig. 28 AIC SEQUENCE screen
4.
Press the displayed setting to be changed. The corresponding setting screen will
be displayed.
WAIT or HOLD
The following screen for time setting will appear.
Fig. 29 A screen for time setting (WAIT TIME)
Item
Settable range
Description
WAIT
0 min to 999 min
Set the waiting time for stabilization after changing gas.
The recommended setting is 3 minutes or longer.
HOLD 0 min to 999 min
Set the calibration validation time (to check the readouts on the recorder after finishing the
calibration).
CAL.
YES/NO
Specify whether or not to perform calibration.
YES: Calibration is performed. NO: Calibration is not performed.